A Comprehensive Guide to Gas Chromatography-Mass Spectrometry (GC-MS)

Updated: Aug 26

1. What is gas chromatography-mass spectrometry?

Gas chromatography-mass spectrometry (GC-MS) combines two independent analytical techniques: gas chromatography (GC) and mass spectrometry (MS). These two techniques complement each other and collaborate on data through an interface. In conventional analytical systems, a gas chromatograph is connected to a mass spectrometer via a heated transfer line, operating in tandem—the chromatograph performs component separation, while the mass spectrometer handles detection and identification. Notably, some miniaturized or portable specialized instruments have integrated the entire GC-MS system into a single device module through integrated design.

Gas chromatography (GC) is an analytical technique based on physical separation. It separates the components of a mixture by analysing the differences in the distribution coefficients of compounds between the stationary phase of a chromatographic column and the mobile phase of a carrier gas. Qualitative and quantitative analysis of these components is performed based on retention times and detector response intensities. However, the output information from traditional GC detectors (such as FID and ECD) has significant limitations: ① The data dimension is limited to two-dimensional parameters (retention time + response value), which cannot provide molecular structure information; ② The identification mechanism relies on comparison of retention times with standards, which lacks the ability to resolve unknown compounds or co-eluting peaks in complex matrices.

To address this technical bottleneck, gas chromatography coupled with mass spectrometry (GC-MS) has become a key solution. This coupled system can operate in two modes: 1) Single mass spectrometry detection mode : The entire column effluent is directed to the mass spectrometer, enabling highly specific detection through characteristic fragment ions; 2) Split detection mode : A microfluidic splitter distributes the post-column effluent to the mass spectrometer and a conventional GC detector (such as an FID/TCD), simultaneously acquiring structural and quantitative data. Automatic matching of the mass fragmentation spectra (such as characteristic ion peaks generated by the EI source) provided by the mass spectrometer with spectral libraries such as those from NIST increases the accuracy of identifying unknown compounds to over 90%.

 

2. Structural composition of GC-MS

GC-MS is a combination of gas chromatography and mass spectrometry, combining the high separation efficiency of chromatography with the high sensitivity detection and structure analysis capabilities of mass spectrometry. Its core process can be divided into three stages:

Gas chromatography separation: The sample is vaporized and carried into the chromatographic column by the carrier gas, and physical separation is achieved based on the difference in the distribution coefficient of the compounds between the stationary phase and the mobile phase.

Mass spectrometry ionization and detection: The separated components enter the ion source of the mass spectrometer (such as electron impact source EI) through a heated transfer line, are ionized into charged ions under a vacuum environment, and then separated by m/z through a mass analyzer (such as a quadrupole or time-of-flight analyzer).

Data acquisition and analysis: The detector records the ion intensities at different m/z values, generating a mass spectrum (reflecting the fragmentation characteristics of the compound) and a total ion current chromatogram (TIC, reflecting the retention time and peak area of each component).

 

Typical GC-MS mass spectra
Typical GC-MS mass spectra

Figure 1. Typical GC-MS mass spectra (green and orange), total ion chromatogram (red), and mass spectrum (blue).

 

1. Gas chromatography part

The basic process of gas chromatograph is shown in Figure 2. It mainly includes the following five systems: carrier gas system, injection system, separation system, temperature control system and detection and recording system.

(1) Carrier gas system: including gas source, gas purification, gas flow rate control and measurement. To obtain pure and stable carrier gas.

(2) Injection system: It includes the injector and the vaporization chamber. The injector is divided into gas injector and liquid injector. The vaporization chamber is a device that instantly vaporizes the liquid sample.

(3) Separation system: It includes chromatographic column, column oven and temperature control device. The components are separated in the chromatographic column according to the difference in distribution coefficient or adsorption coefficient between the mobile phase and the stationary phase.

(4) Temperature control system: controls the temperature of the vaporization chamber, column box and detector.

(5) Detection and recording system: including detectors, amplifiers, recorders, or data processing devices, workstations. The concentration or mass of each component is converted into an electrical signal and recorded.

 

Gas chromatograph
Gas chromatograph

Figure 2 Gas chromatograph

2. Interface

In a GC-MS system, the interface is a key component for enabling the coordinated operation of chromatography and mass spectrometry, and must simultaneously meet the following technical requirements:

Vacuum adaptation: The operating vacuum of the mass spectrometry ion source needs to be maintained at 10⁻³–10⁻⁵ Pa, while the outlet of the gas chromatography column is at normal pressure (approximately 10⁵ Pa). The interface achieves pressure gradient transition through step-by-step pressure reduction to prevent overload of the mass spectrometry vacuum system.

Component enrichment: Carrier gas (such as helium and hydrogen) accounts for more than 99% of the chromatographic effluent. The interface needs to selectively remove the carrier gas to concentrate the analytes 10–100 times before entering the ion source, thereby improving detection sensitivity.

3. Mass spectrometry

The basic components of a mass spectrometer are an ion source, a mass analyzer, and a detector. In a GC-MS system, gaseous molecules separated by gas chromatography are bombarded by the ion source, electrolytically fragmenting into molecular ions, which are then further broken down into fragment ions. Under the combined action of electric and magnetic fields, these molecules are separated by their m/z values and detected, recorded, and organized by the detector, resulting in a mass spectrum, enabling qualitative and quantitative analysis of the sample.

 

3. How does GC-MS work?

1. After the sample is injected into the gas chromatography system manually or through an automatic sample injector (injection accuracy ±0.1 μL), it is transported to the vaporization chamber (Figure 1(3)) by a high-purity carrier gas (helium/hydrogen) [Figure 1(1)]. The liquid sample is instantaneously vaporized in a heating chamber at 200-300 °C (temperature control accuracy ±1°C), and the injection volume is controlled by split/splitless mode: in split mode, more than 90% of the vaporized sample is quickly discharged through the solenoid valve to avoid overloading the chromatographic column due to high concentration components; in splitless mode, the valve is closed for 60-120 seconds to allow the full introduction of trace substances. The vaporized gaseous components are pushed into the chromatographic column by the carrier gas [Figure 1(4)].

2. Sample components (i.e., analytes) are separated based on their interactions with the mobile phase (carrier gas) and the stationary phase. For most compounds, separation is primarily achieved by gas-liquid partition chromatography, where the analytes partition between the carrier gas (mobile phase) and the liquid stationary phase coating the inner wall of the chromatographic column. For more volatile gases (such as permanent gases), separation is achieved by gas-solid adsorption chromatography, where the stationary phase is a solid material with adsorption properties.

GC-MS systems most commonly use capillary columns (also known as open tubular columns), typically with an inner diameter of 0.1-0.25 mm and a length of 10-30 m. These columns primarily feature a wall-coated open-ended (WCOT) structure, where the liquid stationary phase is chemically bonded to the column inner wall, forming a uniform thin film. This design combines high separation efficiency (up to several thousand theoretical plates) with low column bleed, making it particularly suitable for separating complex mixtures. It should be noted that while gas-solid chromatography columns (such as PLOT columns) are still used for specific gas analyses, liquid stationary phases have become mainstream due to their improved reproducibility and separation selectivity.

3. After separation, unless the analytes are isomers, total baseline resolution is not required for GC-MS analysis, and the neutral molecules enter the mass spectrometer via a heated transfer line [Figure 1 (5)].

4. In a mass spectrometer, neutral molecules must first be ionized and converted into charged ions for detection. The most commonly used electron ionization (EI) process is as follows: an electron beam emitted by a heated filament is accelerated to 70 eV and collides with gaseous molecules entering the ion source. When the high-energy electron knocks an electron out of a molecule, a positively charged molecular ion (M⁺·) is generated, which also has free radical properties (hence the name radical cation). Because the electron energy of 70 eV is significantly higher than most chemical bond energies (typically in the range of 3-10 eV), the molecular ion will undergo fragmentation due to the excess energy. This process may involve the following mechanisms:

Homolytic cleavage: chemical bond breaking to produce two free radicals, with the positively charged fragment retaining its charge

Heterolytic cleavage: The breaking of a chemical bond to produce a positive ion and a neutral fragment

Rearrangement reaction : such as McLafferty rearrangement caused by hydrogen atom migration

The secondary ions (fragment ions) produced by fragmentation are always of lower mass than the original molecular ion, and their distribution pattern depends on:

➤ Molecular composition: The type and position of functional groups directly affect bond breaking energy

➤ Molecular structure: Stereochemical effects may inhibit or promote specific fragmentation pathways

➤ Charge localization: Following Stevenson's rule, after ionization, the charge is preferentially retained on atoms with lower ionization potential

➤ Cleavage site selection: weak bonds (such as C—C single bonds adjacent to heteroatoms) are more likely to break

The resulting characteristic fragment ion group constitutes the "fingerprint" mass spectrum of the compound, in which the molecular ion peak (if not completely fragmented) corresponds to the molecular weight of the compound, while the fragment ions provide structural diagnostic information.

5. After ionization is completed, the mass spectrometer separates the ions through a mass analyzer [Figure 1 (8)]. The core principle of this step is to use electromagnetic fields to apply differential motion trajectory control to ions of different m/z, thereby achieving precise sorting based on mass-to-charge ratio. Commonly used mass analyzers (such as quadrupoles, time-of-flight, or ion traps) adjust parameters such as voltage, magnetic field strength, or measure ion flight time, ultimately allowing only ions of a specific m/z to reach the detector. This process provides the basic signal resolution for subsequent ion detection and mass spectrum generation.

The performance and cost of mass spectrometers vary primarily due to the type of mass analyzer used and the corresponding mass resolution. Mass resolution is defined as the instrument's ability to distinguish between two ions of adjacent mass-to-charge ratios, and is typically quantified as the full width at half maximum (FWHM). Instruments can be divided into two categories based on their resolution level:

Unit mass resolution instruments (e.g., quadrupole, linear ion trap): have a resolving power in the range of R = 1000-4000, can distinguish ions with nominal masses (integer mass units) or m/z differences ≥ 0.1, and are suitable for routine qualitative/quantitative analysis;

High-resolution mass spectrometry (HRMS) (such as Orbitrap, FT-ICR): With a resolution of R=50,000-1,000,000, it can distinguish m/z differences as small as 0.0001-0.001, meeting the needs of accurate mass determination and analysis of trace compounds in complex matrices.

The quadrupole mass analyzer is the most commonly used unit mass resolution device. Its core operating principle is to form a high-frequency oscillating electric field by dynamically adjusting the combination of radio frequency voltage (RF) and direct current voltage (DC). Only ions of a specific m/z can maintain a stable oscillation trajectory in this field and pass through the quadrupole. The remaining ions collide with the quadrupole due to excessive amplitude and are annihilated, thus achieving the purpose of ion screening by mass. Quadrupole instruments mainly support two data acquisition modes:

Full scan mode , which acquires all ions within a mass range, is useful for identifying unknowns, method development, and qualitative and quantitative analysis of high-concentration analytes.

Selected ion monitoring (SIM) mode acquires only selected ions representing the target compound and is useful for trace analysis because higher sensitivity can be achieved, but only for the target analyte.

The two modes can be operated alternately through time division (similar to the MRM mode in GC-MS/MS), balancing broad-spectrum screening with highly sensitive quantification in complex matrix analyses. However, it should be noted that the quadrupole scan speed is inversely proportional to the mass range. Setting a wide mass range (e.g., m/z 50-800) may result in loss of signals from low-abundance ions.

6. After being separated by m/z by the mass analyzer, the ion beam finally reaches the ion detector [Figure 1 (9)], which realizes signal conversion and amplification through the following mechanisms:

Signal conversion: Ions impacting the detector surface induce secondary electron emission, converting the ion flow into a measurable current signal. Amplification: Electron multiplier (EM, suitable for instruments with unit mass resolution): A 10⁴-10⁷-fold current gain is achieved through multi-stage dynodes, with a response time of <1 ns; Microchannel plate array (MCP, commonly used in HRMS systems): Based on a honeycomb microporous structure, multi-ion beam signals are synchronously amplified, with a gain of 10³-10⁴ times and excellent spatial resolution. Data acquisition: The amplified analog signal is input into the data system through a high-speed analog-to-digital converter (sampling rate ≥10 GHz) [Figure 1 (10)], and a chromatogram and a mass spectrum for each data point are generated through peak integration and mass-to-charge ratio calibration .


Simplified diagram of a gas chromatograph-mass spectrometer
Simplified diagram of a gas chromatograph-mass spectrometer

Figure 3 Simplified diagram of a gas chromatograph-mass spectrometer

(1) Carrier gas, (2) Autosampler, (3) Gas inlet, (4) Analytical column, (5) Interface, (6) Vacuum, (7) Ion source, (8) Mass analyzer, (9) Ion detector, (10) Computer


4. GC-MS chromatogram

GC-MS data is three-dimensional, as shown in Figure 4. The x-axis shows retention time; the time from sample injection to the end of the GC run. This can also be thought of as the scan number, which is the number of data points acquired by the mass spectrometer during the entire run. The y-axis is the reactivity or intensity measured by the ion detector. The z-axis is the m/z of the ions within the mass range acquired.

 

GC-MS data is three-dimensional
GC-MS data is three-dimensional, showing scan number/retention time, response/intensity, and m/z.

Figure 4 GC-MS data is three-dimensional, showing scan number/retention time, response/intensity, and m/z.

 

A two-dimensional chromatogram, such as the one shown in Figure 5, is generated by summing the abundances of all ions from a single data point and plotting them against retention time (RT)/scan number to create a total ion chromatogram (TIC), which is more comparable to chromatograms produced by a GC detector. However, each data point in a total ion chromatogram is a separate mass spectrum, which can typically be opened in a separate window within the software. In the example in Figure 5, the apex data point of Peak 3 has been opened.

 

Total ion chromatogram (TIC) output from GC-MS
Total ion chromatogram (TIC) output from GC-MS

Figure 5 Total ion chromatogram (TIC) output from GC-MS

 

Figure 6 shows an example mass spectrum of the straight-chain hydrocarbon n-decane. The molecular ion, m/z 142, is visible on the far right. Because decane is a saturated hydrocarbon, the excess energy generated by ionization cannot be delocalized internally, causing most of the molecular ion to fragment, resulting in numerous fragment ions and a low abundance of the molecular ion. As the chain length of the saturated hydrocarbon increases, the abundance of the molecular ion decreases, until no molecular ion is observed in the mass spectrum.

 

However, unsaturated molecular ions, especially those with conjugated double bonds, such as aromatic compounds, fragment less because the excess energy can be internalized more easily. Another phenomenon observed in the mass spectrum of decane is a series of fragment ions at m/z 43, 57, 71, 85, 99, and 113, which differ by 14 in their m/z.

 

These are formed by overlapping cleavage of bonds of consecutive -C 2 H 4 - units and, if the charge is +1, this corresponds to a mass of 28 unified atomic mass units (u), a key feature of the mass spectrum of hydrocarbons. A mass spectrum is a fingerprint of a molecule and, if acquired using the same ionization technique and voltage, can be compared with a library of spectra acquired using the same technique and voltage.

 

The most common commercial spectral libraries are EI spectra generated at 70 eV. Mass spectra can also be interpreted by the mass of the ions, the presence of isotopes, and the loss of fragment ions to determine the molecular formula and structure of the molecule.

 

Example of the mass spectrum of the straight-chain hydrocarbon decane
Example of the mass spectrum of the straight-chain hydrocarbon decane

Example of the mass spectrum of the straight-chain hydrocarbon decane (C 10 H 22 )

 

In GC, retention time is used to identify target analytes, and area is typically used for quantification. For accurate quantification, chromatographic peaks require good chromatographic separation and baseline resolution, as shown by the chromatographic peaks for RT1 and RT2 in Figure 5. For GC-MS, mass spectrometry provides an additional approach, allowing confirmation of target analytes using either the full mass spectrum or the presence and relative proportions of several ions. Quantification using GC-MS data is typically based on the area of a single, unique ion, as it is less likely to include interference from co-occurring peaks than using the area under a TIC peak. Therefore, chromatographic baseline resolution is not required for accurate quantification, as long as a unique ion can be selected that is not present in the blended peaks, allowing spectral resolution of these peaks and baseline-to-baseline integration.

 

 

5. Advantages and Disadvantages of GC-MS

excellent point

Chromatographic detector can be omitted for general applications

GC is used as an injection system to separate the sample to be tested and directly introduce it into the mass spectrometer for detection. It not only meets the requirement of mass spectrometry analysis for sample uniformity, but also eliminates the tedious process of sample preparation and transfer. It not only avoids sample contamination, but also effectively controls the mass spectrometer injection volume, reduces the contamination of the mass spectrometer, and greatly improves the separation, qualitative and quantitative efficiency of the mixture .

High qualitative ability

Identifying components using chromatographic retention time combined with a compound's fingerprint mass spectrum is significantly superior to relying solely on chromatographic retention time. GC-MS qualitative indicators include molecular ions, functional group ions, ion peak intensity ratios, isotope ion peaks, parent and daughter ion mass numbers of ion reactions , and the retention time windows corresponding to total ion current chromatographic peaks, selected ion chromatographic peaks, and selected reaction chromatographic peaks.


Separation of unresolved chromatographic peaks

Using extracted ions, selected ion monitoring or selected reaction monitoring, combined with certain data processing methods (such as AMDIS), chromatographic peaks that have not been separated or are masked by chemical noise on the total ion current chromatogram can be separated.


Improves quantitative analysis accuracy

Quantitative accuracy and qualitative ability can be improved by using isotope dilution and internal standard techniques .


Improve instrument functionality and automate analysis

The multifunctionality of computers makes the structure of instruments simpler, the operation more convenient , and it is easier to automate the analysis work.


No foot

The analysis objects are limited

The analysis objects are limited to samples that can be vaporized and ionized at around 300°C or below . Compounds that are easily decomposed during heating and are too polar, such as organic acids, require esterification derivatization before GC-MS analysis. If the sample cannot be vaporized or esterified, LC-MS or other methods must be used for analysis .

Test conditions are limited

The sample for GC-MS analysis should be an organic solution. Organic matter in solid or aqueous solution cannot generally be measured and needs to be extracted and separated into an organic solution, or thermal cracking or headspace injection technology must be used.


Mass spectrometry applications are limited

Another important shortcoming of mass spectrometry at present is that it cannot distinguish many isomers (especially positional isomers).


 

6. Main Applications of GC-MS

GC-MS, with its high sensitivity, high selectivity, and dual advantages of combining chromatographic separation with mass spectrometry, has become a core tool in modern analytical chemistry. The following detailed discussion covers six key areas: environmental science, food safety, medicine and clinical practice, forensic science and public safety, industrial manufacturing, and life science research , encompassing over 30 specific application scenarios.

1. Environmental monitoring and pollution control

(1) Volatile organic compound (VOCs) detection, such as monitoring of benzene series (such as benzene, toluene, xylene), formaldehyde, chlorinated hydrocarbons and other carcinogens in the atmosphere

(2) Analysis of persistent organic pollutants (POPs), such as ultra-trace detection of dioxins (PCDD/Fs) (detection limit up to ppt level), used to assess emissions from waste incineration plants

(3) Ecotoxicology research: Analyze the accumulation effects of pollutants in organisms (fish, birds) and establish a biomarker database

2. Food Safety and Quality Control

(1) Pesticide residue and veterinary drug detection: for example, rapid screening of pyrethroids and organophosphorus pesticides in fruits and vegetables

(2) Analysis of food additives and illegal additives: such as testing the compliance of sweeteners (saccharin sodium, aspartame) and preservatives (benzoic acid, sorbic acid) in beverages

(3) Analysis of flavor and aroma components: such as qualitative and quantitative analysis of coffee volatile aroma components (pyrazines, furans), and optimization of roasting process

3. Medical diagnosis and drug development

(1) Clinical metabolomics: such as the discovery of VOCs markers (such as pentane and acrolein) in the exhaled breath of cancer patients, assisting non-invasive diagnosis

(2) Pharmacokinetic studies: such as bioavailability assessment of volatile components (e.g., menthol, eucalyptol) in traditional Chinese medicine compounds (3) Drug quality control: such as migration monitoring of plasticizers (DEHP) in injections to ensure the safety of packaging materials

4. Forensic Science and Public Safety

(1) Criminal investigation: such as trace detection of explosives (TNT, RDX), used for counter-terrorism and security screening

(2) Toxic and drug analysis: such as structural analysis and metabolite tracking of new synthetic cathinone drugs (3) Biological evidence identification: such as correlation analysis of characteristic lipid markers in body particulate matter (dander, hair) for suspect identification

5. Industrial manufacturing and energy sectors

(1) Petrochemical analysis: such as detailed spectrum analysis of crude oil components (normal alkanes, isoalkanes) to optimize refining processes

(2) Materials Science: Identification of gases released from electronic components (such as volatile siloxanes) to prevent equipment failures

(3) Energy and Environmental Engineering: such as monitoring the composition of landfill gas (methane, hydrogen sulfide) and optimizing energy recovery

6. Life Sciences and Frontier Research

(1) Metabolomics and systems biology: such as the construction of fingerprints of microbial volatile metabolites (MVOCs) for rapid identification of pathogens

(2) Space and astrochemistry: such as the detection of organic molecules in Martian soil samples, searching for signs of life (3) Doping detection: such as ultra-sensitive detection of anabolic steroids (such as nandrolone, stanozolol) and their metabolites in athletes' urine

Future development trends

Miniaturization and on-site detection: Application of portable GC-MS (such as Torion T-9) in emergency response to achieve real-time monitoring of pollution incidents

High-resolution mass spectrometry: GC-HRMS (e.g., Orbitrap) enhances the identification of unknowns in complex matrices and facilitates non-targeted screening.

AI-assisted analysis: Intelligent matching of machine learning algorithms (such as convolutional neural networks) with mass spectra to shorten data analysis time

Green analytical methods: Rapid GC-MS method development with low solvent consumption, in line with green chemistry principles.


References

McLafferty, F. W., & Tureček, F. (1993). Interpretation of Mass Spectra. University Science Books.


Gross, M. L. (2004). High-Resolution Mass Spectrometry: Principles and Practice. Wiley-Interscience.


de Hoffmann, E., & Stroobant, V. (2007). Mass Spectrometry: Principles and Applications. John Wiley & Sons.


Purnell, R. (2010). Gas Chromatography and Mass Spectrometry: A Practical Guide. Academic Press.


Stull, D. R. (1991). Gas Chromatography: A Practical Approach. Springer.



National Institute of Standards and Technology (NIST). (2020). NIST Mass Spectral Library. Retrieved from NIST website.


Wenzel, T. J., & Smith, R. D. (2018). Applications of GC-MS in Environmental Analysis. Environmental Chemistry Letters, 16(1), 1-20. https://doi.org/10.1007/s10311-017-0665-3


Liu, X., & Wang, Z. (2019). Recent Advances in GC-MS Techniques for Food Safety. Food Control, 98, 193-204. https://doi.org/10.1016/j.foodcont.2018.11.025


Kuo, J. J., & Kim, H. Y. (2021). Clinical Applications of GC-MS in Metabolomics. Journal of Chromatography B, 1168, 122-129. https://doi.org/10.1016/j.jchromb.2021.122129


Skoog, D. A., West, D. M., & Holler, F. J. (2013). Fundamentals of Analytical Chemistry. Cengage Learning.

 

 
 
 

Comments


HOVOGEN (Headquarter)

Dongguan, China

Floor 4, Room 411, No. 10, Keji 2nd Road, Zhongke Innovation Plaza, Songshan Lake High-Tech Industrial Development Zone, Dongguan City, Guangdong, China 523000
Tel/WhatsApp: +86-15916927868 Email: sales@hovogen.com

HOVOGEN (Manufacture Base) Zhuzhou, China

Add:Unit 1208, Building C1, Xinma Jingu Phase III, No. 266 Xinma South Road, Majiahe Street, Tianyuan District, Zhuzhou City, Hunan

HOVOGEN (Office)

HongKong, China

Add:Unit 18, 10/F

Fortune Commercial Building

362 Sha Tsui Road

Tsuen Wan 

Hong Kong

Tel: +852-8402 1602

HOVOGEN (Japan Branch)

Tokyo, Japan

東京都八王子市長沼町104-2ヒルサイドテラス1-4

bottom of page